Teledyne LGA-3500 - Laser Gas Analysis system User Manual

Page 58

Advertising
background image

Operation Model

3500

Teledyne Analytical Instruments

44

• Con. output--20mA: The measured gas concentration output is in

the form of 4-20mA. Con. output--20mA is the gas concentration
value corresponding to 20mA.

Figure 4.20 Concentration Output Menu


The buttons on this menu refer to:

4mA:

Set the corresponding gas concentration (0-
99.9999)% when the gas concentration output signal
is 4mA.

20mA:

Set he corresponding gas concentration (0-99.9999)%
when the gas concentration output signal is 20mA.

4. Purge Parameter Menu

On the Purge Parameter Menu (Figure 4.21), the selections are:

• Pur Len: Set the purging path length (Lf2+Lf1+Lb1+Lb2, Figure

4.22).

• Pur. Temp: Set the purging gas temperature.
• Pur. Con.: Set the concentration of gas under test in the purging

gas.