Fairbanks Series II Bench Scales User Manual
Page 10
07/10
10
50777 - Rev. 6
6. Press the [AutoTare] key to go to the division size mode.
d=0.01
Press the [ON/F] key to select the appropriate division size
d=0.02
d=0.05
d=0.1
d=0.2
d=0.5
7. Press the [AutoTare] key to go to SPAN mode.
12222
(Offset value to be displayed around 5000 ~ 50000 )
8. Press the [ZERO] key to set the display to zero.
0
9. Put on the calibrating weight in lb or kg.
122225
(Span value to be 100000 ~ 250000 at full capacity )
10. Press the [AutoTare] key and the indicator will cycle
600
through a list of permissible weights. When the weight
on the scale is displayed, press the [AutoTare] key.
11. The calibration is now complete, install the JP3 jumper to secure the calibration.
Note: Step 6 is available only in scales with serial number S05159000150 and above.