A. Function Keys

= Select

B. Calibration Procedure

1. Remove the jumper JP3 to calibrate the scale.

2. Press and hold any key while powering ON.

The system goes to Calibration mode.

CAL

Displays:

Definition

3. Press the [AutoTare] key to select the

AUTO ZERO mode

AO 0

None

Press the [ON/ F] key to select the desired choice.

AO 1AZT

0.6d

AO 2AZT

1.0d

AO 3AZT

2.0d

4. Press the [AutoTare] key to step to the

WEIGHING UNITS mode

Un i t.LB

lbs mode

Press the [ON/ F] key to select weighing units in lb or kg.

Un i t.kG

kg mode

5. Press the [AutoTare] key to step to the CAPACITY mode.

C XXXXX

Press the [ON/ F] key to select the appropriate Capacity.

Section 4: Calibration

= [AutoTare] = Enter