A. Function Keys
= Select
B. Calibration Procedure
1. Remove the jumper JP3 to calibrate the scale.
2. Press and hold any key while powering ON.
The system goes to Calibration mode.
CAL
Displays:
Definition
3. Press the [AutoTare] key to select the
AUTO ZERO mode
AO 0
None
Press the [ON/ F] key to select the desired choice.
AO 1AZT
0.6d
AO 2AZT
1.0d
AO 3AZT
2.0d
4. Press the [AutoTare] key to step to the
WEIGHING UNITS mode
Un i t.LB
lbs mode
Press the [ON/ F] key to select weighing units in lb or kg.
Un i t.kG
kg mode
5. Press the [AutoTare] key to step to the CAPACITY mode.
C XXXXX
Press the [ON/ F] key to select the appropriate Capacity.
Section 4: Calibration
= [AutoTare] = Enter