Calibration steps – Fairbanks Aegis Industrial Mild Steel User Manual
Page 14

Section 3: Scale Installation
08/12
14
51102 Rev. 7
Calibration Steps
Adjust the analog interface indicator to the platform.
•
Install all the corners to within
one (1) division of each other at 25% of
rated capacity.
•
Follow the appropriate indicator service manual to ensure a proper calibration.
STEPS
1. Center the four Junction Box Potentiometers by turning the adjustment screw
counter-clock-wise position until a clicking sound is heard, then turning each of
them back clock-wise
ten (10) turns
.
─
Total number of turns is twenty.
2. Identify the platform corner numbers.
3. Place a concentrated weight (
25%
of platform capacity)
on corner #1, then move it to #2, #3 and #4, noting the
displayed reading on each corner.
4. Identify the lowest reading, and then place the
concentrated weight on this corner.
1
2
3
4
50612-1