Installation Requirements and Instructions


PHYSICAL SPACE REQUIREMENTS:


 See Safety Precautions for important information regarding the installation area.

 The unit must be installed on a level concrete pad/floor and protected from the weather and freezing temperatures.

 Approximate overall dimensions/space requirements are listed below. Additional space is required around and

above the equipment for loading and unloading, storage drums, access for maintenance, etc.

Equipment model(s)

Width

Depth

Height

LS-15E/15IIE

51 in. (1.3 m.)

27 in. (0.7 m.)

42 in. (1.1 m.)

LS-15E/15IIE + JetVac

108 in. (2.7 m.)

27 in. (0.7 m.)

48 in. (1.2 m.)

LS-55E/55IIE

75 in. (1.9 m.)

40 in. (1.0 m.)

60 in. (1.5 m.)

LS-55E/55IIE + JetVac

131 in. (3.3 m.)

40 in. (1.0 m.)

60 in. (1.5 m.)


IMPORTANT: If using a Little Cooler closed-loop cooling system with the Little Still, the Little Cooler must be
located outside of the hazardous area. Refer to the Little Cooler Instruction Manual for further details.

ASSEMBLY:

See Parts Location Diagram for additional information.


 Remove the unit from the shipping containers and inspect for damage. If damage is detected, save all packaging

and immediately report any exterior or interior damage to the freight carrier.


 Remove all accessories from the inside of the Still (and the JetVac’s cabinet, if applicable).

 Install the pressure relief valve assembly into the top of the Still’s lid. Face the elbow towards the rear of the unit

(away from the occupied area).


 Attach the supplied grounding cables/clamps to the ground block on the right side of the Still.

 Thread the discharge hose into the discharge pipe on the right side of the Still. Omit this step if installing a JetVac

(see below).

SPECIAL ASSEMBLY FOR LS-15IIE AND LS-55IIE:

“System II” Stills have additional features for

automatic operation. Explanations of these features are discussed in the Instrumentation and Controls section.

 Install the vapor temperature gage into the control panel on the front of the Still (this gage is a delicate instrument

and is packed separately to prevent damage). Attach the white wire to the positive (+) terminal and the red wire to
the negative (-) terminal. Mount the gage using the two nuts provided.