QUAD-4Track User’s Manual Revision 2.1

18

If the control is OFF, then no correction will be done.

Auto Set

Calls the Autoset VI. This VI attempts to find the best range and trigger level for the
signal being measured. The flow is:

1. Set trigger level to 7%

2. Set Range to top Range.

3. Test for triggering.

If no triggers are present, the VI reduces the range to the next lowest and tests for
triggering. It repeats this process until the lowest range is reached. If no triggers are
present, it then reduces the trigger level 1% at a time until either triggering occurs or the
minimum trigger level is reached.

When a trigger is found, the VI returns the discovered setup to the Position Data Display.

Run

Data Collection Control. When active the data is collected for display if the device is
triggered. Data Collection is suspended if another control is activated, and then resumed
when the other control is released.

Indicators

The indicators on the Position Data Display are:

X (mm)
Y (mm)
Total
Quad Energy A
Quad Energy B
Quad Energy C
Quad Energy D
Frequency (Hz)
Triggered
Data Stream On
Wave Comp On
Position Calibrated
Pulse Track Plot


X (mm)