Up17p-6s-w5 – Gentec-EO UP and XLP12 User Manual
Page 16

Version 5.0
Ultra Series UP Instruction Manual
Gentec Electro-Optics Inc. All rights reserved
12
a
The calibrations at 2.1 to 2.5 µm are on special request only.
b
Nominal value, actual value depends on electrical noise in the measurement system.
c
Including linearity with power
d
See graph at the end of this section
Specifications subject to change without notice.
UP17P-6S-W5
UP17P-6S-W5
Effective Aperture Diameter
17 mm
Spectral Range
190 nm
– 10 µm
Calibrated spectral range
a
0.248
– 2.5 µm
Power Noise Level
b
1 mW
Typical Rise time (0-95%)
5 s (1.4 s with anticipation)
Typical sensitivity
0.6 mV/W
Calibration Uncertainty
c
± 2.5 %
Linearity with Power
± 2 %
Repeatability (Precision)
± 0.5 %
Power Resolution
± 0.5 %
Max. Average Power
6 W
Max. Average Power (1 min)
(cooling : minimum 3 min)
7 W
Max. Average Power Density
d
1.064 µm, 10 W CW
100 kW/cm
2
Pulsed Laser Damage Thresholds
1.064 µm, 150 µs, 5 Hz
1.064 µm, 7 ns, 10 Hz
532 nm, 7 ns, 10 Hz
266 nm, 26 ns, 10 Hz
Max. Energy Density
100 J/cm
2
1.1 J/cm
2
1.1 J /cm
2
0.7 J /cm
2
Peak Power Density
667 kW/cm
2
157 MW/cm
2
157 MW/cm
2
27 MW/cm
2
Dimensions (mm)
46(H) x 46(W) x 10.7(D)
Weight (head only)
0.1 kg
Recommended load Impedance
100 kΩ
Output Impedance
N.A.
Linearity vs beam dimension
± 0.5 %
PCB electrical supply
N.A.
Max output signal
N.A.