Operating instructions, Tech tip – Robinair TIF8800X Combustible Gas Detector User Manual

Page 8

Advertising
background image

6

Operating Instructions

Once the battery pack is fully charged, the TIF8800X is ready to use .
1 . Move to a known, non-contaminated atmosphere .
2 . Turn the sensitivity control fully counter-clockwise .
3 . Turn the power switch On . The power light should be illuminated and no

sound should be heard . Once the instrument has warmed up for about 30

seconds, a ticking sound will be heard .

4 . Rotate the sensitivity control to the highest setting that does not cause the

instrument to alarm .

Tech Tip

5 . Search the general area of the suspected leak .

When a detectable compound enters the tip, the

tick rate speeds up . Along with the increased tick

rate, the leak intensity LEDs illuminate from left to

right. The larger the concentration, the more LEDs

will light up . See Figure 4 .

6. In most cases, it is not necessary to adjust the

sensitivity of the instrument . However, if the alarm

sounds before a leak source can be found, it is

likely the air is contaminated with heavy

concentrations of gas. You may desensitize the

instrument by turning the sensitivity knob counter-

clockwise to a lower setting in order to pinpoint the

leak location .

Leak intensity LEDs

Figure 4

Joint compounds used on newly installed

piping may contain combustible solvents

and could cause a false alarm .

Advertising