Automation Command and Query Reference Manual - Processor Reference
Format
Enum
Values
Bathtub
EstTIE
LogEstTIE
LogTIE
TjGaussian
TotalJitter
Frequency
Double
From 100000 to 1e+011 step 1
Range
MaxPopInFit
DoubleLockstep
From 10 to 1e+009 step 20, locked to 1 2 5, fine grain allowed=false, on=false
Range
PercentileUsed
DoubleLockstep
From 0.001 to 20 step 0.001, locked to 1 2 5, fine grain allowed=false, on=false
Range
TransitionDensity
Double
From 0.1 to 1 step 0.01
Range
UseValInput
Bool
UseWeights
Bool
INTEGRAL
app.Math.Fx.OperatorYSetup (Operator = "Integral")
Integral of the linearly rescaled (multiplier and adder) input.
' Visual Basic Script
Set app = CreateObject("LeCroy.XStreamDSO")
' Configure F1 to integrate C1
app.Math.F1.View = True
app.Math.F1.Source1 = "C1"
app.Math.F1.MathMode = "OneOperator"
Example
Adder
Double
AutoFindScale
Bool
FindScale
Action
Multiplier
Double
VerOffset
Double
VerScale
DoubleLockstep