Digital – Teledyne LeCroy WaveRunner Automation Command User Manual
Page 42

Automation Command and Query Reference Manual - Control Reference
PatternState
Enum
Sets/Queries the pattern state for the input channel Cx. Only valid when the trigger mode is set to
'Logic'.
Description
' Visual Basic Script
Set app = CreateObject("LeCroy.XStreamDSO")
' Set the pattern state for channel C1 to low.
app.Acquisition.Trigger.C1.PatternState = "Low"
Example
Values
DontCare
High
Low
Slope
Enum
Sets/Queries the direction of the transition to be used for internal triggering from channel Cx.
Description
' Visual Basic Script
Set app = CreateObject("LeCroy.XStreamDSO")
' Set the direction of the transition as negative
' for triggering on channel C1.
app.Acquisition.Trigger.C1.Slope = "Negative"
Example
Values
Either
Negative
Positive
Window
WindowSize
Double
From 0.02 to 0.205 step 0.0005
Range
DIGITAL
app.Acquisition.Trigger.Digital
Coupling
Enum
InputImpedance
Enum
Level
Double
Level2
Double
PatternState
Enum
Slope
Enum
WindowSize
Double
1-34