Where Ctotal is the combined probe and circuit capacitance and Rtotal is

combined circuit and probe resistance.
For a setup where Ct = 0.9 pF and a source resistance is 250

Ω, the measured

rise time will be 0.495 ns, which will correspond to a bandwidth of 909 MHz,

assuming no inductive loads.

t

rise

= 2.2 x 0.9 x 10-

12 x 250 Ω = 0.495 ns

Parallel combination of 250

Ω and 1 MΩ is still 250 Ω

Example Probe input equivalent circuit. Actual diagrams p.14.

To illustrate the effect of capacitive loading at higher frequencies:
At a frequency of 750 MHz the reactance of the 0.9 pF capacitance is 236

Ω, and

at 1.0 GHz the reactance has been lowered to 177

Ω.

If, at a given frequency, the source impedance is large with respect to the input

impedance, a measurable reduction in the output signal amplitude may occur.

where Zprobe is the probe’s input impedance and Zsource is the source

impedance.
For example: At 750 MHz, where the probe input impedance has reduced to 236

Ω, and a source resistance of 250 Ω the probe output amplitude is reduced to: