Performance verification test record – Teledyne LeCroy ZS2500 User Manual

Page 30

Advertising
background image

ZS Series High-Impedance, Active Probes

20. Remove the banana plug adapter from the DMM and connect the precision

50 Ω terminator to the DMM, making sure that the banana plug side marked

"GROUND" is connected to the LOW or COMMON input of the DMM.

21. After the DMM has stabilized, record the reading to 100 μV resolution as

"Measured Output Voltage, mid range" in the Test record.

22. Calculate the error by dividing the measured mid range output voltage

recorded in step B-21 by the expected mid range output voltage recorded in

step B-19. Subtract 1 from this ratio and multiply by 100% to get the error in

percent.

23. Record the calculated error to two decimal places (±0.xx %) as "Gain Error,

mid range" in the Test record.

24. Verify that the mid range gain error is less than ±1.0%
This completes the Performance Verification of the ZS1000, ZS1500, or ZS2500.

Complete and file the Test Record, as required to support your internal

calibration procedure.
Apply suitable calibration label to the ZS1000, ZS1500, or ZS2500 housing as

required.
NOTE: The function generator used in this Performance Verification Procedure is

used for making relative measurements. The output of the generator is

measured with a DMM or oscilloscope in this procedure. Thus, the generator is

not required to be calibrated.

Performance Verification Test Record

The next page can be used to record the results of measurements made during

the performance verification of the ZS Series Probes. Permission is granted to

photocopy it and record the results on the copy. File the completed record as

required by applicable internal quality procedures. The section in the test record

corresponds to the parameters tested in the performance verification

procedure. The numbers preceding the individual data records correspond to

the steps in the procedure requiring the recording of data.
Results to be recorded in the column labeled Test Result are the actual

specification limit check. The test limits are included in all of these steps. Other

measurements and the results of intermediate calculations that support the limit

check are to be recorded in the column labeled Intermediate Results.

24

924282-00 Rev A