Specifications, Nominal characteristics, Warranted characteristics – Teledyne LeCroy HFP1000 User Manual

Page 11: Typical characteristics

Advertising
background image

Operator’s Manual

922251-00 Rev A

5

Specifications

Nominal Characteristics

Nominal characteristics describe parameters and attributes that are guaranteed by design, but
donot have associated tolerances.

Input Dynamic range

±8 V

Maximum Input Voltage

1

40 V pk

Intended Output Load

50 Ω

Attenuation

÷10

Output Connector

ProBus

Interface

ProBus

Oscilloscope Full Compatibility

Teledyne LeCroy WaveRunner and WavePro
oscilloscopes, and LC series oscilloscopes
with firmware version 8.7.0 or higher.

1

Subject to input voltage vs. frequency derating. See Figure 1.

Warranted Characteristics

Warranted characteristics are parameters with guaranteed performance. Unless otherwise noted,
tests are provided in the Performance Verification Procedure for all warranted specifications.

Low Frequency

±1% plus uncertainty of 50 Ω termination

Attenuation Accuracy

Output Zero

≤ 8 mV, referred to the input

Typical Characteristics

Typical characteristics are parameters with no guaranteed performance. Tests for typical
characteristics are not provided in the Performance Verification Procedure.

Output Zero

≤ 4 mV, referred to the input

Bandwidth (Probe only)

1.5 GHz

Input Capacitance

0.7 pF

DC Input Resistance

100 kΩ