Specifications, Nominal characteristics, Warranted characteristics – Teledyne LeCroy HFP1000 User Manual
Page 11: Typical characteristics
![background image](/manuals/353477/11/background.png)
Operator’s Manual
922251-00 Rev A
5
Specifications
Nominal Characteristics
Nominal characteristics describe parameters and attributes that are guaranteed by design, but
donot have associated tolerances.
Input Dynamic range
±8 V
Maximum Input Voltage
1
40 V pk
Intended Output Load
50 Ω
Attenuation
÷10
Output Connector
ProBus
Interface
ProBus
Oscilloscope Full Compatibility
Teledyne LeCroy WaveRunner and WavePro
oscilloscopes, and LC series oscilloscopes
with firmware version 8.7.0 or higher.
1
Subject to input voltage vs. frequency derating. See Figure 1.
Warranted Characteristics
Warranted characteristics are parameters with guaranteed performance. Unless otherwise noted,
tests are provided in the Performance Verification Procedure for all warranted specifications.
Low Frequency
±1% plus uncertainty of 50 Ω termination
Attenuation Accuracy
Output Zero
≤ 8 mV, referred to the input
Typical Characteristics
Typical characteristics are parameters with no guaranteed performance. Tests for typical
characteristics are not provided in the Performance Verification Procedure.
Output Zero
≤ 4 mV, referred to the input
Bandwidth (Probe only)
1.5 GHz
Input Capacitance
0.7 pF
DC Input Resistance
100 kΩ