Preliminary procedure – Teledyne LeCroy HFP1000 User Manual

Page 36

Advertising
background image

HFP1000 High Frequency Probe

30

922251-00 Rev

A

Table 3 List of Required Equipment

Description

Minimum Requirements

Example Test Equipment

Digital Oscilloscope

ProBus interface

Teledyne LeCroy LT344
Teledyne LeCroy LC584

Digital Multimeter (DMM) with test
probe leads

4.5 digit DC: 0.1% Accuracy AC:
0.1% accuracy

Agilent Technologies 34401A
Fluke 8842A-09

BNC Coaxial Cable (2 ea.)

Male to Male, 50 Ω, 36" Cable

Pomona 2249-C-36
Pomona 5697-36

Calibration Fixture

ProBus Extender Cable

Teledyne LeCroy PROBUS-CF01

Terminator, Precision, BNC

50 Ω ± 0.05%

Teledyne LeCroy TERM-CF01

Preliminary Procedure

1. Remove the two screws that secure the plastic cover on the cable end of the ProBus

interface housing.

2. Gently pull on the probe cable to slide the circuit board assembly from the metal

housing.

3. Connect the HFP1000 probe to the female end of the ProBus extension cable, being

careful to line up all six pins of the probe connector. Connect the male end of the
ProBus extension cable to channel 1 of the oscilloscope.

4. Apply power to the oscilloscope and test equipment.

5. Allow at least 30 minutes warm-up time for the HFP1000 and test equipment before

starting the calibration procedure.