Test equipment required, Operator’s manual, Table 3 list of required equipment – Teledyne LeCroy HFP1500 User Manual

Page 37

Advertising
background image

Operator’s Manual

922252-00 Rev A

31

Test Equipment Required

Table 3 List of Required Equipment lists the test equipment and accessories (or their equivalents)
that are required for complete calibration of the HFP1500 Probe. Specifications given for the test
equipment are the minimum necessary for accurate calibration. All test equipment is assumed to be
correctly calibrated and operating within the specification listed. Detailed operating instructions for
the test equipment are not given in this procedure. Refer to the test equipment manual if more
information is needed

If alternate test equipment is substituted, control settings or calibration equipment setups may
need to be altered.

Table 3 List of Required Equipment

Description Examples

Minimum Requirements

Example Test Equipment

Digital Oscilloscope

ProBus interface

Teledyne LeCroy WavePro960
Teledyne LeCroy LT344

Digital Multimeter (DMM)
with test probe leads

4.5 digit
DC: 0.1% Accuracy
AC: 0.1% Accuracy

Agilent Technologies 34401A
Fluke 8842A-09

Power Supply

0-12 V, settable to 10 mV

HP E3611A

BNC Coaxial Cable (2 ea.)

Male to Male, 50 Ω, 36" Cable

Pomona 2249-C-36
Pomona 5697-36

BNC Tee Connector

Male to Dual Female

Pomona 3285

Calibration Fixture

ProBus Extender Cable

Teledyne LeCroy PROBUS-
CF01

Terminator, Precision, BNC

50 Ω ± 0.05%

Teledyne LeCroy TERM-CF01

Banana Plug Adapter (2 ea.)

Female BNC to Dual Banana
Plug

Pomona 1269

BNC to Mini-grabber

BNC Male to Mini-grabber
Cable, 36"

Pomona 5187-C-36

Advertising