Page 174

164
Merlin II Protocol Analyzer User’s Manual
CATC
SW Version 2.50
Step 2
Enter MerlinII.exe -CLK
The application appears on the screen.
Step 3
Select Setup > Recording Options to open the Recording Options
dialog box.
Step 4
Set Inquiry Timeout to 0.
Step 5
Press the Start Recording button run inquiry for at least 5 minutes.
This will bring the analyzer to its operating temperature.
Step 6
Press Stop to stop the inquiry.
Step 7
Select Setup >Connectors from the menu.
The following window appears:
Step 8
Select the option Output Clock for Calibration.
At this point the clock signal should be directed to the 'TRIG OUT'
BNC connector. The frequency counter should start showing a value
around 26 MHz now.
Step 9
Click on the Clock Cal... button. Please note that this button is
visible only when the Merlin II application is run with the -CLK
argument.