ArbStudio

94

922244-00 Rev A

A

CQUISITION

W

RITE

M

EMORY

M

ODE

When the Digital Pattern Generator works as a sampler, it stores the data it
reads from the input pins to internal memory. This memory can be written
in One Shot or Cyclic modes.

One Shot - The data is written until the end of the memory is
reached.

Cyclic - The memory is written in a cyclic mode and data can be
overwritten when the end of memory is reached.

S

AMPLING

S

ETTINGS

The Sampling Rate Prescaler (SRP) may assume values ranging from 1
(maximum frequency) up to 16777216 (minimum frequency), by multiples
of 2.

The Real Sampling Rate = Sampling Rate / SRP.

For example setting SRP = 20 with a sampling rate of 125 MS/s, the real
sampling rate becomes 6.25MS/s and it is possible to generate digital
waveforms with total length of 167.77216 ms with a time resolution of
160 ns.

Pattern Settings - The Trigger Event Tab

The Trigger Event tab allows setting a trigger condition on Ch (17..0) and on
Vector(16) (Ext Trig) to generate a trigger event similar to a logic analyzer.

M

ASTER

T

RIGGER

E

VENT

(S

LAVE

M

ODE

O

NLY

)

On a slave device, Ch17 can be used as an input to evaluate the trigger
condition (Master Trigger Event disabled) or as a Master Trigger Event.

The device set as the master transmits the Trigger Event by the Expansion
Bus to all slave devices.

If you enable the Master Trigger Event on a slave device, it can also trigger
on a master trigger event condition.