Tracer TP-35000 M-L A03122 User Manual

Page 3

Advertising
background image

3

English

MaximaTP-35000 Series

Ultra-High Intensity UV-A Lamps

U.S. Patent 6,573,665

DANGER

Ultraviolet radiation emitted from this product. Avoid exposure. ALWAYS
WEAR PROTECTIVE CLOTHING. EXPOSURE MAY CAUSE PREMATURE
AGING OF THE SKIN AND CANCER. ALWAYS WEAR PROTECTIVE EYEWEAR;
FAILURE TO DO SO MAY RESULT IN SEVERE BURNS OR LONG TERM
INJURY TO EYE. Never look directly into the lamp. Exposure can cause eye
and skin allergy and allergic reactions. Medications or cosmetics may
increase your sensitivity to ultraviolet radiation. Consult physician before
operating this product if you are using medications or have a history of skin
problems or believe yourself especially sensitive to sunlight.

FOR PROFESSIONAL EXAMINATION USE ONLY

IMPORTANT SAFETY INSTRUCTIONS
1. Read all instructions.
2. Care must be taken, as burns can occur from touching hot parts.
3. Do not operate the lamp with a damaged cord or if the lamp has been damaged.
4. You must wear the TP-9940 fluorescence-enhancing/protective glasses supplied to

protect your eyes and to make leaks clearly visible.

5. Don’t touch the hot filter of the lamp. Never operate the lamp with the UV filter lens

removed.

6. Never use this equipment in any manner not specified in these instructions

because the protection provided by the equipment may be impaired.

SAVE THESE INSTRUCTIONS

OPERATION
1. Turn off the engine before performing any leak detection inspections.
2. Add fluorescent dye to the system. For best results, use OEM-approved Tracerline

Fluoro-Lite® and Dye-Lite® dyes. See the dye instructions for full information.

3. Run the engine and/or system for 5 to 10 minutes to circulate the dye. Turn off

the engine.

4. Connect the plug to a power source that complies with the electrical requirements

specified on the lamp’s label.

5. Push the switch on the ballast. The lamp will achieve maximum output within

seconds.

6. While wearing the TP-9940 glasses, scan the system with the lamp. The leaks will

show best under low light conditions.

7. When the lamp is used intermittently at frequent intervals, leave the lamp burning

continuously rather than turning it on and off repeatedly. This extends bulb life. If the
lamp is turned off, it can be restarted immediately.

8. After the leaks are repaired, clean off the dye residue with GLO-AWAY™ dye

cleaner and reinspect with the lamp.

OPERATOR SERVICING

The following instructions are intended for limited servicing of the lamp. Do not
perform any servicing that is not covered in these instructions.
Return the lamp
to the factory for maintenance not covered in this section.

HOUSING AND FILTER MAINTENANCE

CAUTION: Always shut off the switch on the ballast and unplug the ballast from
the power source before performing any maintenance or servicing. Allow the
lamp to cool before handling.

14

GARANTIE
Die Garantiebestimmungen für diese UV-Lampen befinden sich auf dem Blatt Begrenzte
Garantie", das jedem Gerät separat beiliegt.
HINWEIS: Wenden Sie sich bei Fragen jeglicher Art an die Kundendienstabteilung
(Customer Service Department) von Tracer Products. Die gebührenfreie Nummer
innerhalb der USA und Kanadas lautet: 1-800-641-1133. Beschreiben Sie das Problem im
Detail und geben Sie die Modell- und Seriennummer des Geräts sowie das Kaufdatum
an. Sollte es notwendig sein, das Gerät an das Werk zurückzusenden, wird man Ihnen
Anweisungen für den Versand geben. Falls Sie einen Kostenvoranschlag für nicht unter
die Garantie fallende Arbeit oder andere Leistungen benötigen, erhalten Sie diesen nach
der Begutachtung des Geräts. Arbeiten, die nicht unter die Garantie fallen, werden ohne
Zustimmung des Kunden nicht durchgeführt.

Vorsicht (siehe Begleitunterlagen)

POSTEN TEILE BESCHREIBUNG

3

106081 FILTERHALTER, SILIKON

5

TP-1050 FILTERLINSE
TP-1050D FILTERLINSE (DIFFUNDIEREND)

7 TP-35100 BIRNEN- /REFLEKTOR-SET, 35 W SPOTLICHT

TP-35100/F BIRNEN- /REFLEKTOR-SET, 35 W FLUTLICHT

9

118821 STARTER/FASSUNG

11

PC-0270 LAMPENGEHÄUSE

13

TP-1601A PISTOLENGRIFF

15

EH-0124 ZUGENTLASTUNG (NICHT ABGEBILDET)

17

119799 KABEL, SEKUNDÄR, 2,4 M

19

118820 VORSCHALTGERÄT ELEKTRONISCH

21

100253 WIPPENSCHALTER

23

119797 ABNEHMBARER AC KABELSATZ, 2,4 M, 120 V,100 V

25

WC-0105 ABNEHMBARER AC KABELSATZ, 2,4 M, 230 V

27

119954 GRIFF, VORSCHALTGERÄT

29

MC-0158 BEFESTIGUNGSSTIFT

Advertising