ATRT-03, ATRT-03A, AND ATRT-03B USER’S MANUAL

REV 7

13

2.0 PRE-TEST

SETUP

2.1

ATRT-03 and ATRT-03B Operating Voltages

The ATRT-03 and ATRT-03B’s voltage is preset at the factory and is selectable between 110-120
Vac, 50/60 Hz or 220-240 Vac, 50/60 Hz. The voltage is set by placing jumpers on the reference
transformer as listed in Table 7 and illustrated in Figure 4.

Table 7. Voltage Selection Jumper Settings

Voltage

Selection

Transformer Jumpers

110 – 120 Vac

Pin 1 & 3, Pin 2 & 4

220 – 240 Vac

Pin 2 & 3

Figure 4. Voltage Selection Jumper Location and Settings

2.2 ATRT-03A

Operating

Voltages

The ATRT-03A uses a special switching power supply. The input AC receptacle can accept
voltages that range from 90-264 Vac or 110-240 Vdc. The ATRT-03A can also accept 12 Vdc
from a battery via the front panel jacks.

2.3

LCD Screen Contrast Control

To increase the LCD screen contrast, press and hold the

[PAPER Contrast]

key for two

seconds. Release the button when the desired contrast level has been reached.

To decrease the LCD screen contrast, press and hold the

[PAPER Contrast]

key for two

seconds. Release the button when the desired contrast level has been reached.