Figure 9-12. Typical Slow-Close test report
9.19 Transducer Self-Test
The users can check transducer electronics by connecting transducer to the CT-7500 and
running a “Transducer Self Test”.
Table 9-20. Performing a Transducer Self Test
STEP
OPERATION
ACTION
DISPLAY
1
SELECT DIAGNOSTIC.
PRESS # 4 KEY.
1. SLOW-CLOSE TEST
2. PRINT DATA
3. TEST TRANSDUCER
4. CHECK HOOKUP
2
SELECT TRANSDUCER.
PRESS # 3 KEY.
TRANSDUCER TEST
1= XXX.XX INCHES
2= XXX.XX INCHES
3= XXX.XX INCHES
Note
1. After selecting transducer test mode, move transducer slider and observe display on the CT-
7500 LCD.
2. If slider is moved 1.00”, the CT-7500 LCD should display 1.00”.
3. Press STOP key to return to main menu.