Accu-Chek insulin pump User Manual

Page 175

Advertising
background image

Start-up graph over the stabilization period

The start-up graph shows changes in the flow rate over the stabilization time.

The measurements were made at a medium basal rate of 1.0 U/h according to
IEC 60601-2-24:1998 with a Disetronic Classic infusion set PC 16/110 at room
temperature.

Influence of height on delivery accuracy
The average deviation of the delivered amount (overall mean percentage flow
error) for U100 insulin is

± 10% when your ACCU-CHEK Spirit insulin pump is

located 3.3 feet (1 m) above/below the level of infusion site.

Appendixes

ACCU-CHEK Spirit Insulin Pump

174

Advertising