Accu-Chek insulin pump User Manual

Page 97

Advertising
background image

TBR OVER Alert

At the end of a Temporary Basal Rate, an alert A7: TBR OVER occurs. Press

f

twice to confirm and turn off the alert. See the section “Alert A7: TBR OVER”
(page 153) for more information.

Note:

If an Extended Bolus or a MultiWave Bolus is active at the same time, the
remaining time and amount of the bolus and the amount of the increased
or decreased hourly basal rate are displayed in the RUN screen.

Note:

During the Temporary Basal Rate delivery, the increase or decrease in
percentage, the increased or decreased hourly basal rate (hourly basal rate
plus TBR percentage), and the remaining time are displayed in the RUN
screen. An arrow pointing up symbolizes a Temporary Basal Rate increase;
an arrow pointing down symbolizes a Temporary Basal Rate decrease.

Chapter 4: Boluses and TBRs

ACCU-CHEK Spirit Insulin Pump

96