Current Process Input Procedure

Equipment

Precision current source, 0 to 20 mA range, with
0.01 mA resolution.

Input Setup and Calibration

1.

Connect the correct power supply to terminals 1
and 2 (see Chapter Two).

2.

Connect the current source to terminals 11(-) and
8(+).

3.

Enter 4.00 mA from the current source to the con-
troller. Allow at least 10 seconds to stabilize. Press
the Down Key

¯ until the Input Calibrate 4 mA

parameter,

[``A;4] appears. Press and hold the

ß Key and change [`~no] to [`yES] by press-
ing the Down Key

¯. Release the ß Key and

press the Down Key

¯ to store the 4 mA input and

move to the next parameter.

4.

Enter 16.00 mA from the current source to the con-
troller. Allow at least 10 seconds to stabilize. At the
Input Calibrate 16 mA parameter,

[`A;16] , press

and hold the

ß Key and change [`~no] to

[`yES] by pressing the Down Key ¯. Release the
ß Key and press the Down Key ¯ to store the
16 mA input and move to the next parameter or
press the Infinity Key

ˆ to exit the Factory Page.

5.

Rewire for operation and verify calibration.

Process Output Procedures

Equipment

Precision volt/ammeter with 3.5-digit resolution.

Output 1 Setup and Calibration

1.

Connect the correct power supply to terminals 1
and 2 (see Chapter Two).

Volts

2.

Connect the volt/ammeter to terminals 7 (-) and 6 (+).

3.

Press the Down Key

¯ until the Output 1 Cali-

brate 1 V parameter,

[01;1u] appears. Press and

hold the

ß Key and change [`~no] to [`yES] by

pressing the Down Key

¯. Release the ß Key.

The voltage output value appears in the display.
Press the Up

¿ or Down Key ¯ to adjust the dis-

play value to match the value from the volt/amme-
ter. The unit should stabilize within one second. Re-
peat until the volt/ammeter reads 1.00 V, ±0.1 V.
Press the Down Key

¯ to store the value and move

to the next parameter.

4.

At Output 1 Calibrate 9 V parameter

[01;9u], press

and hold the

ß Key and change [`~no] to

[`yES] by pressing the Down Key ¯. Release the
ß Key. The voltage output value appears in the
display. Press the Up

¿ or Down Key ¯ to adjust

the display value to match the value from the
volt/ammeter. The unit should stabilize within one
second. Repeat until the volt/ammeter reads 9.00 V,
±0.1 V. Press the Down Key

¯ to store the value

and move to the next parameter or press the Infini-
ty Key

ˆ to exit the Factory Page.

5.

Rewire for operation and verify calibration.

Milliamperes

6.

Connect the volt/ammeter to terminals 7 (-) and 5
(+).

7.

Press the Down Key

¯ until the Output 1 Cali-

brate 4 mA parameter,

[01;4a] appears. Press and

hold the

ß Key and change [`~no] to [`yES] by

pressing the Down Key

¯. Release the ß Key.

The current output value appears in the display.
Press the Up

¿ or Down Key ¯ to adjust the dis-

play value to match the value from the volt/amme-
ter. The unit should stabilize within one second. Re-
peat until the volt/ammeter reads 4.00 mA, ±0.1
mA. Press the Down Key

¯ to store the value and

move to the next parameter.

8.

At Output 1 Calibrate 16 mA parameter

[01;16],

press and hold the

ß Key and change [`~no] to

[`yES] by pressing the Down Key ¯. Release the
ß Key. The current output value appears in the
display. Press the Up

¿ or Down Key ¯ to adjust

the display value to match the value from the
volt/ammeter. The unit should stabilize within one
second. Repeat until the volt/ammeter reads 16.00
mA, ±0.1 mA. Press the Down Key

¯ to store the

value and move to the next parameter or press the
Infinity Key

ˆ to exit the Factory Page.

9.

Rewire for operation and verify calibration.

Restoring Factory Calibration

Press the Up

¿ and Down ¯ keys together for six

seconds until

[FaCT] appears in the display.

Press the Down Key

¯ to step through the parame-

ters until the

[rESt] appears.

Press and hold the

ß Key and change [`~no] to

[`yES] by pressing the Down Key ¯. Release the
ß Key and press the Infinity Key ˆ to exit the
Factory Page.