IM 01C22T01-01E

2-12

2. HART COMMUNICATOR OPERATION

F0224.EPS

Set –50, and press ENTER (F4).

EJA:
Engr disp LRV
0.0
0.0

DEL

DEL

ESC

ENTER

EJA:
Engr disp range
1 Engr unit
2 Engr disp LRV
3 Engr disp URV
4 Engr disp point

HELP

SEND

HOME

ENTER

Press to select engr disp URV.

EJA:
Engr disp URV
0.0
0.0

DEL

DEL

ESC

ENTER

Set 50, and press ENTER (F4).
Press SEND (F2) to send data.

HELP

(ENTER)

‘– 5 0’

(ENTER)

‘–5 0’

(SEND)

1. Device setup

4. Detailed setup

4. Display condition

3. Engr disp range

2. Engr disp LRV, 3. Engr disp URV

Example: Set low range value (LRV) to –50 and upper range

value (URV) to 50.

(11) Unit for Displayed Temperature

When the instrument is shipped, the temperature units
are set to

C

(Centigrade). Follow the procedure below

to change this setting.

The unit changed here corresponds the unit for

Snsr

temp

.

F0225.EPS

(ENTER)

EJA:
Snsr temp unit
C
C
F

HELP

SEND

ESC

ENTER

Select F (Fahrenheit), and Press
ENTER (F4).

1. Device setup

4. Detailed setup

1. Sensors

2. Temp sensor

2. Snsr temp unit

Example: Change the unit for the temperature display.

(12) Unit for Displayed Static Pressure

Follow the procedure to change the static pressure unit.

Changing this parameter also changes the unit for the
static pressure display.

F0226.EPS

(ENTER)

EJA:
Static Pres unit
mmH2O
mmH2O
mmHg
psi
bar

HELP

SEND

ESC

ENTER

Select kPa and Press ENTER (F4).
Select SEND (F2) to send the data.

(SEND)

1. Device setup

4. Detailed setup

1. Sensors

3. Static Pres sensor

2. Static Pres unit

Example: Change the static pressure unit from mmH

2

O to

kPa.

inH

2

O

inHg
ftH

2

O

mmH

2

O

mmHg
psi
bar
mbar
g/cm

2

kg/cm

2

Pa
kPa
torr
atm