2) checking with integral indicator, Checking with integral indicator -21, Error messages — dpharp integral indicator – Yokogawa EJA115 User Manual

Page 28

Advertising
background image

IM 01C22T01-01E

2-21

2. HART COMMUNICATOR OPERATION

(2) Checking with Integral Indicator

If an error is detected in the self-diagnostic, an error
number is displayed on the integral indicator. If there is
more than one error, the error number changes at two-
second intervals.

See Table 2.4.1 regarding the error numbers.

F0243.EPS

Figure 2.4.1 Identifying Problems Using the Integral

Indicator

• Error Messages — DPharp Integral Indicator

Integral

Indicator

Display

Description

Cause

Countermeasure

Output Operation

during Error

None

----

GOOD

ERROR

Er. 01

CAP MODULE
FAULT

Er. 02

AMP MODULE
FAULT

Er. 03

OUT OF RANGE

Er. 04

OUT OF SP
RANGE

Er. 05

OVER TEMP
(CAP)

Er. 06

OVER TEMP
(AMP)

Er. 07

OVER OUTPUT

Er. 08

OVER DISPLAY

Er. 09

ILLEGAL LRV

Er. 10

ILLEGAL URV

Er. 11

ILLEGAL SPAN

Er. 12

ZERO ADJ OVER

Static pressure
exceeds specified
range.

Capsule temperature
is outside range
(–50 to 130

°

C).

Amplifier temperature
is outside range
(–50 to 95

°

C).

Output is outside high
or low range limit
value.

Displayed value is
outside high or low
range limit value.

LRV is outside setting
range.

URV is outside setting
range.

SPAN is outside
setting range.

Zero adjustment is too
large.

T0206.EPS

Capsule problem

Amplifier problem

Input is outside
measurement range
limit of capsule.

Displays present
output.

Displays present
output.

Displays present
output.

Outputs high or low
range limit value.

Displays high or low
range limit value.

Holds output
immediately before
error occurrence.

Holds output
immediately before
error occurrence.

Holds output
immediately before
error occurrence.

Displays present
output.

Outputs the signal
according to status of
burnout direction
switch (High or Low).

Outputs the signal
according to status of
burnout direction
switch (High or Low).

Outputs high range
limit value or low
range limit value.

Check line pressure
(static pressure).

Use heat insulation or
make lagging to keep
temperature within
range.

Use heat insulation or
make lagging to keep
temperature within
range.

Check input and range
setting, and change
them as needed.

Check input and
display conditions and
modify them as
needed.

Check LRV and
modify as needed.

Check URV and
modify as needed.

Check SPAN and
change as needed.

Readjust zero point.

Replace capsule
when error keeps
appearing even after
resart.*

1

Replace amplifier.

Check input.

Integral

Indicator

Display

Description

Cause

Countermeasure

Output Operation

during Error

None

----

GOOD

ERROR

Er. 01

CAP MODULE
FAULT

Er. 02

AMP MODULE
FAULT

Er. 03

OUT OF RANGE

Er. 04

OUT OF SP
RANGE

Er. 05

OVER TEMP
(CAP)

Er. 06

OVER TEMP
(AMP)

Er. 07

OVER OUTPUT

Er. 08

OVER DISPLAY

Er. 09

ILLEGAL LRV

Er. 10

ILLEGAL URV

Er. 11

ILLEGAL SPAN

Er. 12

ZERO ADJ OVER

Static pressure
exceeds specified
range.

Capsule temperature
is outside range
(–50 to 130

°

C).

Amplifier temperature
is outside range
(–50 to 95

°

C).

Output is outside high
or low range limit
value.

Displayed value is
outside high or low
range limit value.

LRV is outside setting
range.

URV is outside setting
range.

SPAN is outside
setting range.

Zero adjustment is too
large.

Capsule problem*

1

Amplifier problem

Input is outside
measurement range
limit of capsule.

Displays present
output.

Displays present
output.

Displays present
output.

Outputs high or low
range limit value.

Displays high or low
range limit value.

Holds output
immediately before
error occurrence.

Holds output
immediately before
error occurrence.

Holds output
immediately before
error occurrence.

Displays present
output.

Outputs the signal
according to status of
burnout direction
switch (High or Low).

Outputs the signal
according to status of
burnout direction
switch (High or Low).

Outputs high range
limit value or low
range limit value.

Check line pressure
(static pressure).

Use heat insulation or
make lagging to keep
temperature within
range.

Use heat insulation or
make lagging to keep
temperature within
range.

Check input and range
setting, and change
them as needed.

Check input and
display conditions and
modify them as
needed.

Check LRV and
modify as needed.

Check URV and
modify as needed.

Check SPAN and
change as needed.

Readjust zero point.

Replace capsu

Replace amplifier.

Check input.

*1 : This error code appears at capsule problem or when an illiegal overpressure is applied to the
pressure sensor.
*2 : If the normal pressure is regained, the Er.01 will disappear according to the setting of the parameter
of Auto recover. When the Auto recover is set to ON(default setting), the Er.01 will disappear auto-
matically. When the Auto recover is set to OFF, restart the transmitter to cancel Er.01. If no error
code appears then, perform necessary adjustments such as zero-adjustment to continue the
operation. If the error code still appears. replace the capsule assembly.