2 auto lrv (change low range value), 1 using model 275 — apply values, 2 setting the range using the rangesetting switch – Yokogawa EJA115 User Manual

Page 6: Auto lrv (change low range value) -3, Using model 275 — apply values -3

Advertising
background image

IM 01C22T01-01E

1-3

1. ZERO POINT ADJUSTMENT

1.2 Auto LRV (Change Low

Range Value)

1.2.1 Using Model 275 — Apply Values

Display the Apply Values display, and adjust the zero
point as follows:

1

1. Device setup

3. Basic setup

3. Re-range

2. Apply values

EJA:
WARN-LOOP should be
removed from
automatic control

ABORT

OK

F0105.EPS

Press OK (F4).

(OK)

2

EJA:
Set the:
1 4mA
2 20mA
3 Exit

ABORT

ENTER

Select 4 mA, and press ENTER
(F4)
.

(ENTER)

3

EJA:
Apply New 4ma input

ABORT

OK

Apply a pressure corresponding to
0% of the measurement range.

(OK)

4

EJA:
Current applied
Process value:
0.06 mmH2O
1 Set as 4mA value
2 Read new value
3 Leave as found

ABORT

ENTER

LRV (0%) is read as –0.06 mmH

2

O.

Select Set as 4mA value, and
press ENTER (F4).

(ENTER)

‘Pressure stabilizing’

1.2.2 Setting the Range Using the Range-

setting Switch

With actual pressure(s) being applied to the transmitter,
the range-setting switch (push-button) attached to the
integral indicator plate and the external zero-adjust-
ment screw allow users to change the lower- and
upper-range values for the measurement range (LRV
and URV) without the use of a HART Communicator.
However, a change in the display settings (scale range
and engineering unit) for the integral indicator requires
a HART Communicator.

Follow the procedure below to change the LRV and
URV settings.

Example: Rerange LRV to 0 and HRV to 20 kPa.
1) Connect the transmitter and allow them to warm up

for at least five minutes.

2) Press the range-setting push-button.

The integral indicator then displays

LSET

.

3) Apply a pressure of 0 kPa (atmospheric pressure) to

the high-pressure side of the transmitter.

(Note 1)

4) Turn the external zero-adjustment screw (either in

the output-increase or -decrease direction).
The integral indicator displays the output signal (in
%).

(Note 2)

5) Adjust the output signal to 0% (1 V DC) by

rotating the external zero-adjustment screw. Doing
so completes the LRV setting.

6) Press the range-setting pushbutton.

The integral indicator then displays

USET

.

7) Apply a pressure of 20 kPa to the high-pressure

side of the transmitter.

(Note 1)

8) Turn the external zero-adjustment screw (either in

the output-increase or -decrease direction).
The integral indicator displays the output signal (in
%).

(Note 2)

9) Adjust the output signal to 100% (5 V DC) by

rotating the external zero-adjustment screw. Doing
so completes the URV setting.

10)Press the range-setting pushbutton.

The transmitter then switches back to the normal
operation mode while maintaining the measurement
range at 0 to 20 kPa.

Note 1: Wait until the pressure inside the pressure-detector section

has stabilized before proceeding to the next step.

Note 2: If the pressure applied to the transmitter exceeds the

previous LRV (or URV) , the integral indicator may display
error number “

Er.07

” (In this case, the output signal percent

and “

Er.07

” are displayed alternately every two seconds).

Although “Er.07” is displayed, there is no need to worry and
you may proceed to the next step. However, should any
other error number be displayed, take the appropriate
measure in reference to Section 2.4 Self-Diagnostics,
“Errors Messages.”

Advertising
This manual is related to the following products: