Installation, 1 mounting, 2 mounting the diaphragm seals – Yokogawa EJA115 User Manual

Page 17: 1 mounting 3.2, Mounting the diaphragm seals, Important, 2 mounting the diaphragm seals important

Advertising
background image

<3. Installation>

15

IM 01C22A01-01E

3. Installation

IMPORTANT

• When welding piping during construction, take

care not to allow welding currents to fl ow through
the transmitter.

• Do not step on this instrument after installation.
• For EJA430A, the atmospheric opening is located

on the low pressure side cover fl ange. For
EJA530A with Measurement span code A, B, and
C, the pipe is attached for the opening. These
openings must not face upward.

F00301.ai

Pipe

(Open to atmosphere)

Zero-adjustment screw

Measurment span code

Figure 3.1

EJA530A Horizontal Mounting
Position

3.1 Mounting

■ The transmitter can be mounted on a nominal

50 mm (2-inch) pipe using the mounting bracket
supplied, as shown in Figure 3.2 and 3.3. Tighten the
(four) bolts that hold the transmitter with a torque of
approximately 39 N·m {4 kgf·m}.

F0302.ai

Mounting bracket

Transmitter
mounting bolt

Figure 3.2

Transmitter Mounting
(Horizontal Impulse Piping Type)

F0303.ai

Process connector upside

Process connector downside

Transmitter
mounting bolt

Mounting
bracket

Figure 3.3

Transmitter Mounting
(Vertical Impulse Piping Type)

U-bolt (S)

F0304.ai

U-bolt nut (S)

Plate

Adapter

U-bolt nut (L)

Mounting bracket

50 mm(2-inch) pipe

U-bolt (L)

Figure 3.4

EJA510A and EJA530A Mounting

3.2 Mounting the Diaphragm

Seals

IMPORTANT

• Please use a gasket which has a bigger inside

diameter than that of gasket facing (ød) on
diaphragm seal. In case a gasket which has a
smaller inside diameter than that of gasket facing
is used, it may cause an error as the gasket
prevents diaphragm from working correctly

• During the diaphragm seal installation, ensure as

far as possible that no seal liquid head is applied to
the diaphragm seals.

• Exercise care so as not to damage diaphragm

surfaces. Since the diaphragm protrudes approx.
1mm from the fl ange surface, placing the
diaphragm seals with their diaphragm surfaces
facing downward may damage the diaphragm
surfaces.

• Do not sharply bend or twist capillary tube or apply

excessive stress to them.

• Never loosen the four screws securing the cover

fl

ange or the screws at the joints between the

capillary tube and cover fl anges (if the seal liquid
leaks, the transmitter cannot be used).

Installation

3

Advertising
This manual is related to the following products: