4 iecex certification, 4 iecex certification -9, 6) name plate – Yokogawa EJA115 User Manual

Page 17

Advertising
background image

IM 01C22T02-01E

2-9

2. HANDLING CAUTION

(6) Name Plate

Name plate

Tag plate for intrinsically safe type

Tag plate for flameproof type

F0298.EPS

No. KEMA 02ATEX1344 X
EEx ia C T4

Ui

ϭ

17.5V Ii

ϭ

360mA Pi

ϭ

2.52W Ci

ϭ

1.76nF Li

ϭ

0

or Ui

ϭ

24.0V Ii

ϭ

250mA Pi

ϭ

1.2W Ci

ϭ

1.76nF Li

ϭ

0

EEx ia B T4

Ui

ϭ

17.5V Ii

ϭ

380mA Pi

ϭ

5.32W Ci

ϭ

1.76nF Li

ϭ

0

ENCLOSURE:IP67 Tamb –40 TO 60

°

C PROCESS TEMP. 120

°

C

1

KS25

MODEL: Specified model code.
STYLE: Style code.
SUFFIX: Specified suffix code.
SUPPLY: Supply voltage.
OUTPUT: Output signal.
MWP: Maximum working pressure.
CAL RNG: Specified calibration range.
DISP MODE: Specified display mode.
OUTPUT MODE: Specified output mode.
NO.: Serial number and year of production

*1

.

TOKYO 180-8750 JAPAN:

The manufacturer name and the address

*2

.

*1: The first digit in the final three numbers of the

serial number appearing after “NO.” on the name
plate indicates the year of production. The follow-
ing is an example of a serial number for a product
that was produced in 2001:

The year 2001

12A819857 132

*2: “180-8750” is the zip code for the following

address.

2-9-32 Nakacho, Musashino-shi, Tokyo Japan

2.1.4 IECEx Certification

a.

IECEx Flameproof Type

Caution for IECEx flameproof type.

Note 1. Model EJA Series differential, gauge, and

absolute pressure transmitters with optional code /
SF25 are applicable for use in hazardous locations:

• No. IECEx KEM 06.0005
• Applicable Standard: IEC60079-0:2004,
IEC60079-1:2003
• Type of Protection and Marking Code:
Ex d IIC T6...T4
• Enclosure: IP67
• Maximum Process Temperature: 120

°

C (T4),

100

°

C (T5), 85

°

C (T6)

• Ambient Temperature: –40 to 75

°

C (T4), –40 to

80

°

C (T5), –40 to 75

°

C (T6)

• Supply Voltage: 42 V dc max.
• Output Signal: 4 to 20 mA dc

Note 2. Wiring

• In hazardous locations, the cable entry devices shall

be of a certified flameproof type, suitable for the
conditions of use and correctly installed.

• Unused apertures shall be closed with suitable

flameproof certified blanking elements. (The plug
attached is certificated as the flame proof IP67 as a
part of this apparatus.)

• In case of ANSI 1/2 NPT plug, ANSI hexagonal

wrench should be applied to screw in.

Note 3. Operation

• WARNING:
AFTER DE-ENERGIZING, DELAY 10 MINUTES

BEFORE OPENING.

• WARNING:
WHEN AMBIENT TEMPERATURE

70

°

C,

USE THE HEAT-RESISTING CABLES

90

°

C.

• Take care not to generate mechanical sparking

when accessing to the instrument and peripheral
devices in a hazardous location.

Note 4. Maintenance and Repair

• The instrument modification or parts replacement

by other than authorized representative of
Yokogawa Electric Corporation is prohibited and
will void IECEx Certification.

Advertising