2 csa certification, 3 cenelec atex (kema) certification – Yokogawa EJA115 User Manual

Page 13

Advertising
background image

IM 01C22T02-01E

2-5

2. HANDLING CAUTION

2: Installation should be in accordance with National

Electrical Code (ANSI/NFPA 70) Sections 504, 505
and Local Electrical Code.

3: The configuration of Associated Apparatus must be

Factory Mutual Research Approved.

4: Associated Apparatus manufacturer’s installation

drawing must be followed when installing this
equipment.

5: No revision to drawing without prior Factory Mutual

Research Approval.

6: Terminator and supply unit must be FM approved.
7: Installation requirements;

Vmax

м Voc or Vt

Ca

м Ci + Ccable

La

м

Li + Lcable

2.1.2 CSA Certification

Caution for CSA Explosionproof type

Note 1. EJA Series differential, gauge, and absolute

pressure transmitter with optional code /CF15
are applicable for use in hazardous locations:
• Applicable standard: C22.2 No.0, No.0.4,

No.25, No.30, No.94, No.142, No.1010.1

• Certificate: 1010820
• Explosionproof for Class I, Division 1,

Groups B, C and D.

• Dust-ignitionproof for Class II/III, Division

1, Groups E, F and G.

• Encl “Type 4X”
• Temperature Class:

T6

T5

T4

• Process Temperature: 85

°

C

100

°

C 120

°

C

• Ambient Temperature: –40 to 80

°

C

• Supply Voltage: 32 V dc max.
• Current Draw: 16.5 mA dc

Note 2. Wiring

• All wiring shall comply with Canadian

Electrical Code Part I and Local Electrical
Codes.

• In hazardous location, wiring shall be in

conduit as shown in the figure.

• CAUTION: SEAL ALL CONDUITS

WITHIN 50 cm OF THE ENCLO-
SURE.
UN SCELLEMENT DOIT ÊTRE
INSTALLÉ À MOINS DE 50 cm DU
BÎTIER.

• When installed in Division 2, “SEALS NOT

REQUIRED.”

Note 3. Operation

• Keep strictly the “CAUTION” on the label

attached on the transmitter.

CAUTION: OPEN CIRCUIT BEFORE

REMOVING COVER.
OUVRIR LE CIRCUIT AVANT
D´NLEVER LE COUVERCLE.

• Take care not to generate mechanical spark

when access to the instrument and peripheral
devices in hazardous location.

Note 4. Maintenance and Repair

• The instrument modification or parts

replacement by other than authorized
representative of Yokogawa Electric Corpo-
ration and Yokogawa Corporation of
America is prohibited and will void Cana-
dian Standards Explosionproof Certification.

Non-hazardous
Location
Equipment

32 V DC Max.
15 mA DC
Output current

Non-Hazardous

Locations

Hazardous Locations Division 1

Non-Hazardous

Locations

Hazardous Locations Division 2

50 cm Max.

Sealing Fitting

Conduit

EJA Series

Non-hazardous
Location
Equipment

32 V DC Max.
15 mA DC
Output current

Sealing Fitting

EJA Series

F0201.EPS

2.1.3 CENELEC ATEX (KEMA) Certifica-

tion

(1) Technical Data

a. CENELEC ATEX (KEMA) Intrinsically Safe

Type

Caution for CENELEC ATEX (KEMA) Intrinsically
safe Type.

Note 1. EJA Series differential, gauge, and absolute

pressure transmitters with optional code /KS25
for potentially explosive atmospheres:
• No. KEMA 02ATEX1344 X
• Applicable standard: EN50014:1997,

EN50020:1994, EN50284:1999

• Type of Protection and Marking Code: EEx

ia IIC T4

• Temperature Class: T4
• Enclosure: IP67

Advertising