A6.7 troubleshooting, A-35 – Yokogawa EJA115 User Manual

Page 73

Advertising
background image

IM 01C22T02-01E

A-35

APPENDIX 6. SOFTWARE DOWNLOAD

A6.7 Troubleshooting

For error messages appearing in the download tool, see also the User’s Manual of download tool.

Table 2.

Actions after Software Update

Symptom

Cause

Remedy

An error occurs before starting a

download, disabling the

download.

The selected download file is not for the

selected field device.

Check SOFTDWN_ERROR in the resource

block and obtain the correct file.

An error occurs after starting a

download, disabling the

download.

You attempted to update the device revision

by downloading a file which is not an original

file.

Check SOFTDWN_ERROR in the resource

block and obtain the original file.

The selected field device does not support

software downloading.

Check whether the option code /EE is

included in the model and suffix codes of the

device.

The voltage on the fieldbus segment falls

below the specified limit (9 volts).

Check the capacity of the field bus power

supply used and the voltage at the terminal.

There was an error in a checksum or the

number of transmission bytes.

Check SOFTDWN_ERROR in the resource

block and obtain the correct file.

The download takes far longer

than expected or fails frequently.

The fieldbus segment is noisy.

An error occurs after activation.

Check the noise on the fieldbus segment.

Transient error caused by the internal

resetting of the field device

Check whether communication with the field

device has recovered after a while.

The new software does not take

effect after the activation.

The file of the current revision was

downloaded.

Obtain the correct file.

Failure of the memory in field device, etc.

Check SOFTDWN_ERROR in the resource

block, and re-try downloading.

If fails, place a service call.

TA0602.EPS

The download tool does not allow download

with same software revision.

Check the setting of the download tool.

A6.8 Resource Block’s Parameters Relating to Software Download

Table 3.

Additional Parameters of Resource Block

TA0603.EPS

Relative

Index

Index

Parameter Name

Default

(Factory Set)

Write

Mode

Description

SOFTDWN_PROTECT

1, 0, 0, 0, 0, 0,

0, 0, 0

SOFTDWN_ACT_AREA

Defines whether to accept software downloads.

0x01: Unprotected

0x02: Protected

Selects the software download method.

0x01: Standard

Indicates the number of times the internal

FlashROM was erased.

Indicates the ROM number of the currently

working FlashROM.

0: FlashROM #0 working

1: FlashROM #1 working

Indicates the software module revision.

Indicates the error during a software download.

See Table 4.

1057

1055

1056

SOFTDWN_MOD_REV

1054

1053

0

SOFTDWN_FORMAT

SOFTDWN_COUNT

0x01

0x01

0

56

57

55

54

53

1058

SOFTDWN_ERROR

0

58

Advertising