Yokogawa AXFA14G/C User Manual
Page 12
Advertising

IM 1E6K2-E
4-19-4
4. OPERATION
4.7.3
Meter Factor Setting
The meter factors can be set using three AXFA11 converter keys or HHT.
The standard dual frequency excitation requires two meter factors such as "C21 Low
MF" and "C22 High MF". In case of calibration, two meter factors set to 1.0000.
The enhanced dual frequency excitation requires two meter factors such as "C23 Low
MF (EDF)" and "C24 High MF (EDF)". In case of calibration, two meter foctors set to
1.0000.
For details, refer to the AXFA11 User’s Manual IM 01E20C01-01E.
Advertising
This manual is related to the following products: