General, 1 standard specifications – Yokogawa AXFA14G/C User Manual

Page 3

Advertising
background image

IM 1E6K2-E

3-1

3. GENERAL

3. GENERAL

The AM012 Calibrator is a standard signal generator used to check or calibrate the
ADMAG AXF Magnetic Flowmeter (AXFA11, AXFA14, AXF Integral Flowmeter),
AM11 Magnetic Flow Converter, ADMAG AE Magnetic Flowmeter, ADMAG CA
Magnetic Flowmeter, AM12 Magnetic Flow Converter, AM100A/B or AM200A/B
Magnetic Flowmeter, YMA11 Magnetic Flow Converter and Compact YEWMAG
Magnetic Flowmeter (CYM for short), styles A and B.(Style A or style B is indicated on
the name plate: style A as *A and B as *B.)

The excitation load is incorporated in the calibrator, so the converter can be checked or

calibrated without using a flow detector.

Span can be set from 0 to 10.99 m/s. The output signal related to the span can be set

to 0, 25, 50, 75 and 100% using the output (%) knob.

3.1

Standard Specifications

AM012
Case

: Material

: Unsaturated polyester resin

Dimensions : 110mm (H)

×

235mm (W)

×

364mm (D)

Color

: Light grayish green

Wiring cable

: Special cables supplied are:

AM012 power cord, signal and excitation cables
CYM signal cable and excitation cable (optional)

Range setting

: 0 to 10.99 m/s (minimum scale 0.01 m/s) with the span-

setting dial.
CHECK, 0, 25, 50, 75, 100% with output (%) knob.

Accuracy

:

±

0.04% of set value (span 0.1 m/s or greater)

+0.04 mm/s (span less than 0.1 m/s)

Power supply requirements : 90 to 132 V AC or 180 to 264 V AC, 47 to 63 Hz
Ambient temperature limits : 0 to 40

°

C

Weight

: 4.2kg

Power cord plug

: Three types of plugs are available (select any one of the

following three):
(–1)

For 100 V AC (for domestic use)

(–2)

For 100 V AC (UL Standard)

(–3)

For 200 V AC (CEE Standard)

(–3) 200 V series

(conforms to CEE standard)

(–2) 100 V series

(conforms to UL standard)

(–1) 100 V series

(for domestic use)

Figure 3.1

Power Cord Plugs