Appendix – Yokogawa HART Communicator YHC5150X User Manual

Page 43

Advertising
background image

37

APPENDIX


PRODUCT SPECIFICATIONS

BASE UNIT:
11.9" L, 5.6” W (at display), 2.50” W (at handle), 1.9” D.
PC/ABS alloy plastic case with shock absorbing bumpers, rated IP51, 2.1 lbs including battery. Passes one meter drop
test onto concrete


DISPLAY:
4.3-inch widescreen backlit TFT color touch screen display with 480 x 272 WQVGA pixel resolution.

POWER:
Li-ion battery pack, 3.6 volts nominal.
1.5A recharging cradle. 100-240Vac, 50/60Hz input.

BATTERY LIFE (Approximate):
20 hours typical use

TEMPERATURE SPECIFICATIONS:
Communicator:
Storage: -40° F to 140° F (-40° C to 60° C)
Operating: 14° F to 122° F (-10° C to 50° C)
Charging: 50° F to 104° F (10° C to 40° C)

CONNECTIONS:
Standard banana jacks (¾” center) for HART communication.
Communication (serial cable to PC port) / Battery charging. Connection via recharging cradle.

ATEX Intrinsically Safe II 1 G Demko 13 ATEX 1115457

IECEx Intrinsically Safe Ex ia IIC T4 Ga IECEx UL 13.0004

c

UL

us

Listed

Intrinsically Safe

US

Class I, Div. 1, Groups A, B, C, D: T4

Class I, Zone 0, AEx ia IIC T4

Canada

Ex ia IIC T4


-10º C < Ta < +50º C

Entity parameters:

Pi = 1.25W, Ii = 200mA, Ci = 0, Li=0, Vi=30V, Um (USB) = 5V, Um

(battery) = 5V




MEMORY:
Dedicated µSD flash memory card for YHC5150X Series application software, DD and configuration storage.

ε

Χ

CE compliance

Advertising