5.3 Insulation

The heat exchanger and top flange section of the YARS MUST always be insulated by at least two inches (obtained from

a local supplier). This will reduce the influences of ambient air temperature fluctuations and cause the heat exchanger to

operate more efficiently. Insulation should be installed at site after installation and leak testing and should be configured for

possible future maintenance and access.

It is generally required to have the filter section insulated only when ambient temperatures fall below 32

o

F (0

o

C). Insulation

should conform to appropriate site standards. This insulation may not be required during hot summer months however

only on-site experience can determine the exact insulation requirements for the filter section.