General specifications fu20 – Yokogawa FU20/PH20 All-in-One pH/ORP Sensor User Manual

Page 4

Advertising
background image

IM 12B6J3-02E-E

2

2. GENERAL SPECIFICATIONS FU20

2-1. Measuring elements

: pH glass electrode

: Silver chloride reference

: Solid platinum electrode

: Pt1000 temperature sensor.

2-2. Construction materials
Body

: PPS 40GF (Ryton with glass filling)

Earthing pin

: Solid Platinum

O-ring

: Viton

Reference junction

: Porous PTFE

Cable

: Coaxial with 4 extra leads

2-3. Functional specifications (at 25°C)
Isothermal point

: pH 7

Glass impedance

- Dome shape

: 250 to 500 M

Ω

- Flat Surface

: 400 to 1000 M

Ω

Reference system

: Ag/AgCl with saturated KCI

Junction resistance

: 1.0 to 10 k

Ω

Liguid outled

: non-flow double junction

Temperature element

: Pt1000 to IEC 751

Asymmetry potential

: 8 ± 15 mV

Slope

: > 96 % (of theoretical value)

NOTE: ThetemperaturesensorincludedintheFU20isintendedtoprovideindicationand

cellcompensation.Theconstructionhasnotbeentestedtothepressurevessel
standardsrequiredforplanttemperaturecontrol.

2-4. Dynamic specifications
Response time pH step (7 to 4)

: < 15 sec for 90%

Response time temp step (10°C)

- Dome shape

: < 3 min for 90%

- Flat Surface

: < 6 min for 90%

Stabilisation time (0.02 pH unit/ 10 s)

: < 2 minutes

2-5. Operating range
pH

: 0 to 14

Temperature

: -10 to 105 °C (14 to 221 °F)

Pressure

: 0 to 10 bar ( 0 to 145 PSIG)

Conductivity

: > 50 µS/cm

Storage temperature

: -10 to 50 °C (14 to 122 °F)

NOTE:

Specificationsshouldnotbeconsideredinisolation.ForexamplethepHrangecan
be2-12pH,wherethemeasurementisatelevatedtemperatures.Foradviceabout
specificapplicationspleasecontactyoulocalsalesoffice.