4. WIRING

(pH) A

(Ref.)B

(pH shield) C

(ORP/LE)

D

(Temp)E

(Temp)F

VP-connector

4.1. Cable marking and pin allocation.

FU20 cable

Wire color

WU10-V

cable

Wire color

Signal description

Cable wires

(&EXA terminal)

number

VP connector

Pin

pH

Clear

15

A

Clear

pH-Shield

Brown

16

C

Black

Reference

Black

13

B

Brown

ORP/Liquid earth

Yellow

14

D

Yellow

Temp-1

Red

11

E

Red

Temp-2

Blue

12

F

Blue

4.2. Instrument connection of sensor as function of measurement

Function pH

pH&ORP pH&rH

ORP

pH compen-

Signal descr.

sated ORP

pH

15

15

15

-

13

Shield of pH

16

16

16

-

17

REF

13

13

13

13

-

ORP/Liquid Earth

14

14

14

15

15

TEMP

11

11

11

11

11

TEMP(2)

12

12

12

12

12

Jumper

-

-

-

13-14

14-15

The settings for the pH converters must be set according to the instructions in the
Instruction Manual of the converter. In the following table these settings are summarized.

(pH) Clear A

(Ref.) Brown B

(ORP/LE) Yellow D

(Temp-1) Red

F

(Temp-2) Blue

E

(pH shield) Black C

FU20

WU10

MODEL CODE

Serie nummer

Red (11), Temp-1

ø6.5

Blue (12), Temp-2

Black (13), Reference

Clear (15), pH

Brown (16), pH shield

Yellow (14), ORP/LE

MODEL CODE

Serie nummer

ш6.5

Advertising