General specifications – Yokogawa SC25V User Manual

Page 5

Advertising
background image

5

IM12B6J1-40E-E

2. GENERAL SPECIFICATIONS

2.1 Measuring elements
pH glass electrode
Silver/Silver Chloride reference
Titanium liquid earth
Pt1000 temperature sensor.

2.2 Construction materials
Wetted materials:
O-ring

: Viton

Earth Ring

: Titanium

Reference Junction : Porous PTFE
Measuring Sensor : Glass
Body Sensor

: Glass, PEEK

2.3 Functional specifications
Isothermal point

: pH 7

Glass impedance G-Glass

: 225 M

Ω

Glass impedance L-Glass

: 600 M

Ω

Reference system : Double junction, Ag/

AgCl with saturated KCI,
including Ag+ ion trap

Junction resistance : 0.2 to 10 k

Ω

Temp. element

: Pt1000 to IEC 751

Asymmetry potential : 8 ±15 mV
Slope

: > 96 %

(of theoretical value)

Sensor length

: 120 mm and 225 mm

Note: The temperature sensor included in the

SC25V is intended to provide indication
and cell compensation. The construction
has not been tested to the pressure
vessel standards required for plant
temperature control.

2.4 Operating range
pH

: 0 to 14

Temperature

Type G-glass

: -10 to 80°C

(14 to 176°F)

Type L-glass

: +15 to 130°C

(59 to 266°F)

Pressure

: 0 to 10 bar (0 to 145 psi)*

Conductivity

: > 10 μS/cm

Storage temp.

: -10 to 50°C (14 to 122°F)

* Damaging the screw thread might influence

the max process pressure.

Note: Specifications should not be considered

in isolation. For example the pH range
can be 2-12pH, where the measurement
is at elevated temperatures. For advice
about specific applications please contact
your local sales office.

2.5 Dynamic specifications
Response time pH step (7 to 4)

: < 5 sec for 90%

Response time temp step (25°C)

: < 90 sec for 90%

Stabilization time (0.02 pH unit/10 s)

: < 30 seconds

Advertising