Wiring to the ph analyzer – Yokogawa SC25V User Manual

Page 9

Advertising
background image

9

IM12B6J1-40E-E

4. WIRING TO THE PH ANALYZER

4.1 Cable marking and pin allocation

Signal description

VP connector

Cable Wire Colour

(Using WU10-V-S)

pH

A

Transparent (core coax)

pH-shield

C

Black

Reference

B

Brown (shield coax)

Liquid earth

D

Yellow (cable shield)

Temp -1

E

Red

Temp -2

F

Blue

4.2 Instrument connection of sensor

Instrument Terminal for Measurement

pH

pH&ORP

ORP only

pH-compensated

Signal description

pH&rH

ORP

pH

15

-

13

pH-shield

16

-

17

Reference

13

13

-

Liquid earth / (ORP)

14

15

15

Temp -1

11

11

11

Temp -2

12

12

12

Jumper

-

13-14

14-15

The settings for the pH instruments should follow the instruction Manual of the instrument
The following tables summarize these settings:

4.3 Instrument connection settings for PH402G and PH202G(S)

Function

pH

pH compensated

Setting

(default)

pH&ORP

pH&rH

ORP

ORP

code 01

0(pH)

0(pH)

0(pH)

1(ORP)

1(ORP)

code 02

0(off)

1(ORP)

2(rH)

0(off)

0(off)

code 03

1.1.1

1.1.1

1.1.1

0.0.1

0.0.1

code 04

0.0.1

0.0.1

0.0.1

0.0.1

1.1.1

code 31

0

2

2

Jumper input 1

no

no

no

yes

yes

Jumper input 2

yes

yes

yes

yes

no

4.4 Instrument connection settings for PH150, PH450 and FLXA21

Function

pH

pH compensated

Setting

(default)

pH&ORP

pH&rH

ORP

ORP

sensor setup

pH

pH + ORP

pH + ORP

ORP

ORP

measurement setup

pH + ORP

pH+rH

Impedance settings

input 1

High

High

High

Low

Low

input 2

Low

Low

Low

Low

High

jumper input 1

no

no

no

yes

yes

jumper input 2

yes

yes

yes

yes

no

Advertising