Model ph87 retraction sensor assembly, Sensor replacement instructions – Yokogawa pH/ORP Retractable PH97 User Manual

Page 14

Advertising
background image

14

IM M1263JL-A

Sensor replacement Instructions

BEFOrE SErVICING rEmOVE ThE mODEL Ph87 SENSOr FrOm SErVICE FOLLOwING ThE STEP-By-STEP
INSTruCTIONS [A] AND [D] IN OrDEr AS ShOwN BELOw:

12

Model PH87 Retraction Sensor Assembly

Sensor Replacement Instructions

TO DISASSEMBLE THE MODEL PH87 SENSOR WITH A SENSOR MOUNTED JUNCTION BOX,

FOLLOW THE STEP-BY-STEP INSTRUCTIONS [A] THROUGH [D] IN ORDER AS SHOWN BELOW:

[Step A]. REMOVE

JUNCTION BOX COVER AND DISCON-

NECT ALL WIRES FROM TERMINALS.

[Step C]. WHILE FIRMLY GRIPING

SHEATH PUSH ON COUPLER TUBE -

JUNCTION BOX ASSEMBLY TO

UNSEAT SENSOR O-RINGS

[Step B]. WHILE FIRMLY GRIPING

THE SENSOR AND THE SHEATH AT

THE SAME TIME UNTHREAD THEM

FROM THE COUPLER ASSEMBLY BUT

DO NOT COMPLETELY DISENGAGE

THREADS.

[Step D]. COMPLETELY UNTHREAD

SENSOR FROM COUPLER TUBE

ASSEMBLY.

CONTINUE WITH SERVICE

INSTRUCTIONS AT STEP "N".

[STEP A]. REMOVE
JUNCTION BOX COVER AND DISCONNECT
ALL WIRES FROM TERMINALS.

[STEP B]. WHILE FIRMLY GRIPING
THE SENSOR AND THE SHEATH AT
THE SAME TIME UNTHREAD THEM
FROM THE COUPLER ASSEMBLY BUT
DO NOT COMPLETELY DISENGAGE
THREADS.

[STEP D]. COMPLETELY UNTHREAD
SENSOR FROM COUPLER TUBE
ASSEMBLY.

CONTINUE WITH SERVICE
INSTRUCTIONS AT STEP “N”.

[STEP C]. WHILE FIRMLY GRIPING SHEATH
PUSH ON COUPLER TUBE JUNCTION BOX
ASSEMBLY TO UNSEAT SENSOR O-RINGS