Sensor Installation Overview

NOTE: THE FOLLOWING INFORMATION IS INTENDED TO PROVIDE THE USER WITH A PREVIEW
OF WHAT WILL BE REQUIRED TO INSTALL THE MODEL PH87 SENSOR ASSEMBLY. PLEASE
REVIEW THE DIAGRAM AND GUIDELINES BEFORE PROCEEDING.

Typical Sensor Installation Diagram

Shown with Optional /B Ball Valve

Sensor Installation Guidelines:

[1]. Install the model PH97 sensor in pipe or vessel between 15° and 90° from horizontal.
[2]. Connect flex conduit with adequate service loop to junction box.
[3]. Connect the sensor cable to the terminals inside the junction box.
[4]. Keep cable run between junction box and pH / ORP instrument as short as possible.

Sensor Installation Overview

Sensor Installation Guidelines:
[1]. Install the model PH97 sensor in pipe or vessel between 15˚ and 90˚ from horizontal.
[2]. Connect flex conduit with adequate service loop to junction box.
[3]. Connect the sensor cable to the terminals inside the junction box.
[4]. Keep cable run between junction box and pH / ORP instrument as short as possible.