Yokogawa FU24F User Manual

Page 11

Advertising
background image

IM 12B6J8-01E-E

11

FM

Certificate no.

: 3046277

IS, Class I Div. 1, GP A, B, C, D T3…T6

Electrical data

: For sensor input circuits (by connector), connected to a

FM approved intrinsically safe apparatus meeting the entity

parameters of the SENCOM™ sensor:

Ui = 6.1 V; Ii = 230 mA; Pi = 1.2 W; Li = 4 μH; Ci = 30 μF

or

FM approved intrinsically safe Yokogawa transmitter Model

FLXA21 series.

Ambient temperature: T6 for Tamb. -40 °C to +60 °C

T5 for Tamb. -40 °C to +75 °C

T4 for Tamb. -40 °C to +85 °C

T3 for Tamb. -40 °C to +85 °C

Note: Intrinsically safe when connected as per Control Drawing FF1-K1226QT (see Fig 3)

Control Drawing FM

The FU24F SENCOM

TM

sensor shall be

installed to a FM approved intrinsically safe

apparatus meeting the entity parameters

of the sensor as shown in the table as

maximum values, or to a FM certified

intrinsically safe Yokogawa transmitter

Model FLXA21 series.

When installing this equipment, follow the

manufacturer’s control drawing. Installation

should be in accordance with ANSI/ISA

RP 12.06.01 “Installation of Intrinsically

Safe Systems for Hazardous (Classified)

Locations” and the National Electrical Code

(ANSI/NFPA 70).

To prevent ignition of

flammable or combustible

atmospheres, disconnect

power before servicing or

read, understand and adhere

to the manufacturer’s live

maintenance procedures.

81 82 83 84 85 86 87

+ - G

Hazardous

location

Transmitter

WU11 cable

terminals 82-84,

86-87.

Maximum cable

length: 100 m.

Cable diameter:

5.8mm

Sensor

Ui

6.1 V

Ii

230 mA

Pi

1.2 W

Li

4 µH

Ci

30 µF

Maximum values

Fig 3: FF1-K1226QT Control Drawing FM

FU24F

Model:

NPT

Dome shape

FSM Flat surface

IS, CL I Div.1, GP A, B, C, D T3/T4/T5/T6

Ta 85 ºC/85 ºC/75 ºC/60 ºC

Advertising