General specifications – Yokogawa FU24F User Manual

Page 7

Advertising
background image

IM 12B6J8-01E-E

7

2. GENERAL SPECIFICATIONS

2.1 Measuring elements

pH glass electrode

Ag/AgCl reference

Solid Platinum electrode

Pt1000 temperature sensor

2.2 Wetted parts

Sensor body

: PPS 40GF (Ryton

TM

with glass filling)

Measuring sensor

: G-glass

LE glass tube

: AR-glass

Reference junction

: Porous PTFE

Earth pin

: Solid Platinum

O-ring

: Viton

Bellow system

: Viton

2.3 Functional specifications (at 25°C)

Measuring system

Isothermal point

: pH 7

Reference system

: Ag/AgCl with saturated KCl

Glass impedance

- Dome shape

: 200 MΩ nominal

- Flat surface

: 700 MΩ nominal

Liquid junction

: Non-flow double junction

Junction resistance

: 1 to 15 kΩ

Temperature element

: Pt1000 to IEC 751

Asymmetry potential(zero) : 8 ± 15 mV

Slope

: > 96 % (of theoretical value)

Note: The FU24F temperature sensor is designed for cell compensation and for indication.

It is NOT designed for process temperature control.

2.4 Dynamic specifications

Startup time sensor

: < 60 sec.

Response time pH

: t

90

< 15 sec. (for 7 to 4 pH step at 25 ºC)

Response time temperature

- Dome shape

: t

90

< 1 min. (for 10 °C step)

- Flat surface

: t

90

< 4 min. (for 10 °C step)

Stabilization time pH

: < 2 min. (for 0.02 pH unit during 10 sec.)

Advertising