9

2.8 Regulatory standards

CE

: Decision 768/2008/EC

- ATEX

: Directive 94/9/EC, as amended by Regulation (EC) no.

1882/2003

-

Certificate no.

: DEKRA 11ATEX0064 X

II 1 G Ex ia IIC T3...T6 Ga

-

Electrical data

: For sensor input circuits (by connector) connected to a certified

intrinsically safe circuit with the following maximum values

:

Ui = 6.1 V; Ii = 230 mA; Pi = 1.2 W; Li = 4 µH; Ci = 30 µF

::

or

:

Certified intrinsically safe Yokogawa transmitter

Model FLXA21 series.

-

Special

-

conditions (X)

: T6 for Tamb. -40 °C to +60 °C

T5 for Tamb. -40 °C to +75 °C

T4 for Tamb. -40 °C to +110 °C

T3 for Tamb. -40 °C to +125 °C

: Electrostatic charges on the sensor enclosure shall be avoided.

- Pressure

: Directive 97/23/EC, as amended by Regulation (EC) no.

1882/2003

-

Applying article

: 3.3 (Sound Engineering Practice)

: Damaging the screw thread of the sensor might influence the

maximum process pressure.

- EMC

: Directive 2004/108/EC

IEC 61326-1: 2006 Class A (control and laboratory use)

IEC 61326-2-3: 2006 (use in industrial locations)

- Low Voltage

: Directive 2006/95/EC

:

Sensor contains glass parts which if broken can cause cutting

injuries.

- WEEE

: Directive 2012/19/EU

- RoHS

: Directive 2011/65/EU

IECEx

Applying standards : IEC 60079-0: 2007

IEC 60079-11: 2006

IEC 60079-26: 2006

Certificate no.

: IECEx DEK 11.0065X

Ex ia IIC T3...T6 Ga