Accuracy and display resolution – Yokogawa Button Operated DX2000 User Manual

Page 11

Advertising
background image

10

3. Accuracy and Display Resolution

3.1. Measurement and Display Accuracy (Digital Display)

Computing method of the display value accuracy

¾

Input

voltage

X

¾

Lower limit of display span VL

Upper limit of display span VU

¾

Scale lower limit

SL

Scale upper limit

SU

¾

Digital display value

Y

1) Converting equation for the input voltage (setting)

• Y = 10

(SU-SL)*(X-VL)/(VU-VL)+SL

2) Measurement accuracy of input voltage X (uses the negative side due to the

LOG characteristic)
• X

err

=X – (measurement accuracy of the voltage range)

3) Hardware error of the display value

• Y’= Y – Y

err

• The difference from the true value Y is hardware error Y’ when display

value Y

err

corresponds to input voltage X

err

.

4) Display accuracy of the mantissa

• Display accuracy = Y’* 1.1 + 1 digit

Advertising