Yokogawa Removable Chassis DX1000N User Manual

Page 12

Advertising
background image

2-3

Troubleshooting

SM 04L41B01-01E

1

2

3

4

5

6

7

2.1 Failure Analysis Flow Chart

Is the input wiring correct?

Is a constant ambient

temperature maintained?

Have appropriate measures

against noise been taken?

Are the range, span, and other

input settings appropriate?

Does the input meet the

input specifications?

Problems related to measurement

YES

NO

YES

NO

YES

NO

YES

NO

YES

NO

Ensure that the input

specifications are met.

Set them correctly.

• Enter range, span, and other

input settings that are

appropriate.

• When using the calibration

correction function, check the

settings for the calibration

correction.

• Keep input wires and cables

away from noise sources.

• Be sure to ground the recorder.

• Ground the measured object

properly.

• For thermocouple measurement,

insulate points of contact

between the measured object and

thermocouple.

• Use shielded cables for input

wiring.

• Change the integral time setting

of the A/D converter.

• Use moving average on the input.

• Attach input terminal covers.

• To avoid temperature changes in

the vicinity of input terminals,

keep fan ventilation holes clear,

and keep hot objects away.

• Ensure correct polarity when

wiring.

• Check that terminal wires are

securely connected.

• Do not ground RTDs.

• Replace broken thermocouples.

• Do not connect other instruments

with thermocouples in parallel.

If you must connect it in parallel,

consider using a double element

thermocouple, refraining from

using burnout detection, or

implementing other measures.

• Use one earth connection when

connecting the DX to another

instrument.

Are the RJC

settings and TC, RTD type

settings correct?

See No. 6 of the checklist

in section 2.2.

YES

NO

Are other instruments

connected in parallel?

NO

YES

Advertising
This manual is related to the following products: