Chapter 1 principles of operation, 1 principles of operation, 1 principles of operation -1 – Yokogawa Removable Chassis DX1000N User Manual

Page 8: Block diagram

Advertising
background image

1-1

SM 04L41B01-01E

Principles of Operation

1

2

3

4

5

6

7

Chapter 1

Principles of Operation

1.1 Principles of Operation

The following explains the principles of operation of the DX.

Block Diagram

Data storage functions

Input section

Calculation function

Alarm

Communication function

(optional)

(optional)

Input terminal

Scanner

A/D

CPU

External storage

Internal memory

Display

Remote control

Relay contact

(optional)

(optional)

Pulse input

Transmitter power supply (optional)

USB port

(optional)

Key

Easy text entry function

Input Terminal Section

For connecting the measurement input signal wires. The terminals can be used for DC
voltage, thermocouple, resistance temperature measurement, and contact input.
• The reference junction compensation circuit for thermocouples is built in.
• Proven transistor method for measuring the temperature of the terminals.
• Metal core construction used in the internal printed circuit boards to equalize

temperature.

• Input terminals removable.

Scanner Section

Switches the measured input signal (channel). The input signal switching section uses
highly reliable solid state relays (SSR).

A/D Conversion Section

Converts analog signals to digital signals. It is a PWM (pulse width modulation) type
converter. A/D converter calibration data is stored in an EEPROM.
For the correspondence between input channels and A/D converters, see “Measurement
Accuracy” on page 3-6.