4-4

SM 04L41B01-01E

7.

Press the

End #4 soft key.

DX1000/DX1000N

DX2000

Saving Adjusted Values

8.

Use the

left or right arrow key to choose Yes (save calibration values) or No (do

not save), then press the

DISP/ENTER key.

DX1000/DX1000N

DX2000

9.

Repeat steps

2

8

to adjust all

A/D No.

When finished, press the

ESC key.

Concluding Adjustments

10.

Turn OFF the power switch.

Note

After adjusting the DX1000N, turn ON the power switch then close the bezel. If the power
switch is OFF when you close the bezel, you will have to open it again the next time you use
the instrument.

4.2 Adjusting the A/D Converters