Menu tab, Communication > pop3 details, Math – Yokogawa Button Operated MV2000 User Manual

Page 48: Report, Alarm > basic settings, Alarm > switch, relay, Alarm > hysteresis

Advertising
background image

Communication > POP3 Details

Setting

Selectable Range or Choices

Default Value

POP3 Details

Send delay [second]

0 to 10 s

2

POP3 Login

PLAIN/APOP

PLAIN

Math

Setting

Selectable Range or Choices

Default Value

Math

Value on Error

+Over/-Over

+Over

Value on Overflow > SUM, AVE

Error/Skip/Limit

Skip

Value on Overflow > MAX, MIN, P-P

Over/Skip

Over

Report

Setting

Selectable Range or Choices

Default Value

Report

Report select > 1

Max/Min/Ave/Sum/Inst

Ave

Report select > 2

Off/Max/Min/Ave/Sum/Inst

Max

Report select > 3

Off/Max/Min/Ave/Sum/Inst

Min

Report select > 4

Off/Max/Min/Ave/Sum/Inst

Sum

File type

Separate/Combine

Separate

End

Setting

Selectable Range or Choices

Default Value

Do you want to store and make the new
settings take effect?

Yes/No/Cancel

Menu Tab

Alarm > Basic settings

Setting

Selectable Range or Choices

Default Value

Basic setting

Reflash

On/Off

Off

Rate of change > Decrease

1 to 32

1

Rate of change > Increase

1 to 32

1

Indicator

Hold/Nonhold

Nonhold

Alarm > Switch, Relay

Setting

Selectable Range or Choices

Default Value

Internal Switch

AND

None/S01/S01-S02/…/S01-S29/S01-S30

None

Relay

AND

None/I01/I01-I02/…/I01-I06

None

Action

Energize/De-energize

Energize

Hold

Hold/Nonhold

Nonhold

Relay Action on ACK

Normal/Reset

Normal

Alarm > Hysteresis

Setting

Selectable Range or Choices

Default Value

Hysteresis

Meas CH > High/Low

Numeric value (0.0 to 5.0)

0.5

Meas CH > Delta High/Low

Numeric value (0.0 to 5.0)

0.0

Math CH > High/Low

Numeric value (0.0 to 5.0)

0.0