Yokogawa Standalone MW100 User Manual

Page 27

Advertising
background image

27

2.12.

Changing to Measurement Mode


Disconnect the MX118 from the PC before switching to Measurement mode.

1. Turn OFF the power to the MW100.
2. Turn OFF dip switches 1 and 3–7 on the front of the MX118.

Dip switch number 2 is not used for Measurement mode, and is therefore ignored when in
Measurement mode.
Dip switch number 8 is used for the terminator setting. Set as needed.


4. Turn ON the power to the MW100.

When operating in Measurement mode, the MX118’s STATUS LED lights.

ON

1

2

3
4
5
6
7
8

ON

1

2

3
4
5
6
7
8

When termnator is ON

No. 1: OFF
No. 2: Ignored
No. 3–No. 7: OFF
No. 8: ON

When terminator is OFF

No. 1: OFF
No. 2: Ignored
No. 3–No. 8: OFF