Opposing two-gauge method, Opposing two-gauge three-wire method, Four-gauge method – Yokogawa PC-Based MX100 User Manual

Page 7: Wiring procedure

Advertising
background image

7

IM MX100-72E

• Opposing Two-Gauge Method

Rg1

Rg2

Setting switch

ON

OFF

No.1
No.2
No.3
No.4
No.5

A(+V)

B( L)

C(-V)

D( H)

-B12, -B35

-NDI

No.1

ON

No.2

OFF

No.3

ON

No.4

OFF

No.5

ON

Rg1

Rg2

Bridge head
(701955 or 701956)

Setting switch

R: fixed resistance
r: resistance value of lead wire
Rg: resistance value of strain gauge
e: output voltage from bridge
E: voltage applied to bridge

SW1

ON

SW2

OFF

SW3

ON

SW4

ON

SW5

OFF

1 2 3 4 5

SW

Rg1

Rg2

1
2
3
4
5
6
7
8

ON

OFF

R

R

Rg2

Rg1

E

e

• Opposing Two-Gauge Three-Wire Method

Cannot be connected.
Use -NDI.

Rg1

Rg2

-B12, -B35

-NDI

Bridge head
(701955 or 701956)

Setting switch

R: fixed resistance
r: resistance value of lead wire
Rg: resistance value of strain gauge
e: output voltage from bridge
E: voltage applied to bridge

SW1

OFF

SW2

OFF

SW3
OFF

SW4

ON

SW5
OFF

1 2 3 4 5

SW

Rg1

Rg2

1
2
3
4
5
6
7
8

ON

OFF

e

R

R

Rg2

Rg1

E

r

r

r

r

r

r

• Four-Gauge Method

Rg1, Rg3

Rg2, Rg4

Rg1

Rg3

Rg1

Rg3

Rg4

Rg2

Rg1, Rg2

Rg3, Rg4

Setting switch

ON

OFF

No.1
No.2
No.3
No.4
No.5

A(+V)

B( L)

C(-V)

D( H)

-B12, -B35

-NDI

No.1

OFF

No.2

OFF

No.3

OFF

No.4

OFF

No.5

ON

Rg1

Rg3

Rg2

Rg4

Bridge head
(701955 or 701956)

Setting switch

R: fixed resistance
r: resistance value of lead

wire

Rg: resistance value of strain

gauge

e: output voltage from bridge
E: voltage applied to bridge

SW1

ON

SW2

OFF

SW3

ON

SW4

ON

SW5

OFF

1 2 3 4 5

SW

Rg1

Rg2

Rg3

Rg4

1
2
3
4
5
6
7
8

ON

OFF

Rg4

Rg1

Rg2

Rg3

E

e

5. Wiring Procedure

Advertising
This manual is related to the following products: